Measuring Dyed Fiber Color with MSFU-K Microscope Spectrophotometer: Methodology Validation

https://doi.org/10.30764/1819-2785-2018-13-2-71-80

Abstract

This work is part of a series of efforts towards validation of methods used in forensic fiber analysis. These efforts address current needs for accreditation of forensic laboratories and quality control in operations.

The qualitative testing methodology consists of obtaining absorption spectra with the microscope spectrophotometer MSFU-K and comparing the spectral characteristics of color in fiber samples. The expert determines whether the textile fibers submitted for analysis match in color or not, depending on the results of spectral comparison.

The proposed validation experiment algorithm is designed for evaluating uncertainty in optical density measurements and the level of expert competence.

In this case uncertainty corresponds to reproducibility standard deviation. To evaluate uncertainty, two operators took readings of absorption spectra of dyed fibers independently in the course of three days, and measured optical density at maximum and minimum absorption wavelengths. To evaluate repeatability, 5 spectra were obtained in a row on each of the three days.

The testing was conducted using three samples of polyacrylonitrile (PAN) fibers. Key characteristic points in the samples’ absorption spectra covered a wide range of wavelengths in the visible spectrum. Measurements were taken using the MSFU-K microspectrophotometer, which consists of a microscope with a spectrophotometric add-on unit.

Statistical analysis of measurement data demonstrated uncertainty levels between 7,1 % and 22,1 %. Uncertainty values below 30 % are indicative of quantitative measurements and insignificant variance of optical density values, which corresponds to high reproducibility of spectra and allows the expert to make statistically reliable match/non-match conclusions on the color of compared fibers.

Expert competence was assessed based on «blind» test results. The experts had to determine which of the three samples were colored with the same dye. Each of the two experts was provided with 3 visually identical samples that were colored with different dyes. The experts were asked to distinguish between fibers treated with the same dye. When analyzing obtained spectra, both experts correctly identified same-color fibers based on matching color spectral characteristics.

Positive validation results suggest that the MSFU-K microscope spectrophotometer can be successfully used in forensic fiber analysis for measuring the color of dyed fibers. 

About the Authors

Galina I. Bebeshko
The Russian Federal Centre of Forensic Science of the Ministry of Justice of the Russian Federation
Russian Federation

Doctor of Engineering, Principal Researcher at the Department of Innovations in the Practice of Forensic Science,

Moscow 109028



Irina P. Lyubetskaya
The Russian Federal Centre of Forensic Science of the Ministry of Justice of the Russian Federation
Russian Federation

Leading Researcher at the Department of Innovations in the Practice of Forensic Science,

Moscow 109028



Lidiya P. Brunova
The Russian Federal Centre of Forensic Science of the Ministry of Justice of the Russian Federation
Russian Federation

Lead State Forensic Examiner at the Laboratory of Forensic Fiber Analysis,

Moscow 109028



Marina A. Khanukaeva
The Russian Federal Centre of Forensic Science of the Ministry of Justice of the Russian Federation
Russian Federation

Senior State Forensic Examiner at the Laboratory of Forensic Fiber Analysis,

Moscow 109028



Georgii G. Omel'yanyuk
The Russian Federal Centre of Forensic Science of the Ministry of Justice of the Russian Federation; Peoples' Friendship University of Russia (RUDN University)
Russian Federation

Doctor of Law, Associate Professor, Deputy Director, Moscow 109028;

Professor of the Department of Forensic Operations, Moscow 117198



References

1. Smirnova S.A., Usov A.I., Omeljanuk G.G., Bebeshko G.I., Korol S.G. Practice of Accreditation of Forensic Laboratories of the Ministry of Justice of Russia on Compliance with ISO/MEK 17025. Theory and Practice of Forensic Science. 2012. No 2 (22). С. 40–56. (In Russ.).

2. Fedyanina N.V., Brunova L.P., Lizaeva M.V., Kononets V.A., Nikolaeva E.I. Methodological guidelines for microspeсtrophotometric analysis of single dyed fibers using the MSFU-K microspectrophotometer. Moscow: RFCFS, 2012. 43 p. (In Russ.).

3. Smirnova S., Omelyanyuk G., Bebeshko G., Yudin N. The experience of validation measurement method “The definition of benzo(a)pyrene concentration in the objects of soil and geological origin by means of HPLC fluorimetry detecting” method conducting soil forensic expert examination. Theory and Practice of Forensic Science. 2012. No 3 (27). P. 78–91. (In Russ.).

4. Doerffel K. Statistics in analytical chemistry. Moscow: Mir, 1994. 268 p. (In Russ.).


Review

For citations:


Bebeshko G.I., Lyubetskaya I.P., Brunova L.P., Khanukaeva M.A., Omel'yanyuk G.G. Measuring Dyed Fiber Color with MSFU-K Microscope Spectrophotometer: Methodology Validation. Theory and Practice of Forensic Science. 2018;13(2):71-80. (In Russ.) https://doi.org/10.30764/1819-2785-2018-13-2-71-80

Views: 939


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1819-2785 (Print)
ISSN 2587-7275 (Online)